Medical Physics, MRI Safety and Body Jewelry

Critical numbers

Clinical MRI Field Strengths1.5 T and 3.0 T (standard); 7.0 T (research)
ASTM F2503 MR Safe DefinitionNon-conducting, non-magnetic, no hazard in any MR environment [1]
ASTM F2503 MR Conditional DefinitionSafe only under defined conditions (field, SAR, gradient, location) [1]
Titanium (ASTM F136) χ~1.8 × 10⁻⁴ (paramagnetic, negligible force at clinical fields)
Niobium (ASTM B392) χ~2.4 × 10⁻⁴ (paramagnetic, negligible force at clinical fields)
316LVM Steel (solution-annealed) χ<0.01 (weakly paramagnetic, process-dependent)
316LVM Steel (cold-worked) χUp to 0.05 (ferromagnetic, treat as MR Unsafe)
RF Heating, Closed LoopsRises with loop size, frequency and conductivity
Whole-Body SAR Limit (IEC 60601-2-33, normal mode)2 W/kg averaged over 6 minutes
Gradient Slew Rate (dB/dt)Up to 200 T/s (induces currents in closed conductive loops)
RF Frequency (1.5 T)~64 MHz (¹H Larmor frequency)
RF Frequency (3.0 T)~128 MHz (¹H Larmor frequency)